QUALITY CONTROL AND QUALITY ASSURANCE OF DENTAL RADIOGRAPH
1. DENTAL SYSTEM CONSTANCY TEST
-Steps:
Daily: If manual processing, replenish following the chemistry manufacturer guidelines.
Weekly: Damp mop darkroom floor. Clean counters, cabinets, and anywhere else dust may
-To
assure the processing and dental intraoral x-ray unit are operating
consistently.
2. PROCESOR QC (SENSITOMETRY)
- To determine if film
processor is working optimally in facilities producing radiographs other than
intraoral.
- It should be:
± 0.15 optical density
of the mid-density step
± 0.15 optical density
of the density difference
+ 0.03 optical density
of the base + fog level
-Steps:
1. Expose the control
film with the sensitometer and develop
2. Determine the average
optical density of the mid-density step and record on Form
B1 in
Appendix B.
3. Determine the average
optical density difference and record.
4. Measure the
background optical density (base + fog) and record.
5. Verify that the
measured values are within the suggested performance criteria.
3. DAILY AND WEEKLY DARKROOM QC
-Keep
the darkroom clean and processing optimized.
-Steps:
Daily:If manual
processing, developer temperature must be measured with non-mercury
thermometer for
correlation with the time-temperature chart.
If auto processing,
measure the temperature with a non-mercury thermometer to
verify that the
developer is operating within the temperature range established by the
manufacturer, and that
the display, if applicable, is accurate. It may not be necessary
to
physically measure the temperature daily if the processor passes the daily QC
test (1).
Daily: If manual processing, replenish following the chemistry manufacturer guidelines.
Replace rinse
water.
If auto
processing, follow the processor manufacturer recommendations regarding
replenishment.
Daily: Clean processor feed tray and counter top.
Daily: Clean processor feed tray and counter top.
Weekly: Damp mop darkroom floor. Clean counters, cabinets, and anywhere else dust may
accumulate. Clean film hangers.
4. VIEWBOXES
-To ensure viewboxes are
clean and light levels are kept consistent.
- Suggested Criteria:
Viewbox lights are the
same “color” and luminance, and viewbox surfaces are kept clean.
- Steps:
1. Clean surface of
viewbox.
2. If a bulb or tube
fails, it is best to replace all of them.
3. Record results on the
Monthly QC Checklist (Form 3)
5. VISUAL CHECKLIST
-To assure that all components of the radiographic x-ray
system (indicator lights, displays,
and mechanical locks and
detents) are working properly and that the mechanical rigidity
and stability of the
equipment is optimum.
-Suggested Criteria:
Each of the items listed in the
QC Visual Checklist (Form 2)should pass or receive
a check mark.
Items not passing the visual
check should be replaced or corrected as soon as possible.
6. REPEAT ANALYSIS
-To identify ways to minimize
patient exposure and reduce costs by addressing higher
than normal repeat rates.
- Suggested criteria:
The criteria associated
with repeating a film is subjective. There is no good way to
determine what the
repeat rate should be. Each facility should decide on its own, but
should strive for a
repeat rate of no greater than 5 to 7 %.
7. DENTAL TUBE HEAD AND BOOM STABILITY
TEST
-To
assure the intraoral x-ray tube remains in the correct location.
-Suggested Criteria
: Tube does not drift out of position or
oscillate.
-Steps:
1. Position the tube as
you would with a patient in the chair.
2. Watch for any
movement after you have positioned it.
3. Check all typical
tube positions.
4. Record your results
on the Quarterly QC Checklist (Form 3).
8. FILM AND CHEMICAL STORAGE
-To assure film and
chemistry quality are maintained and inventory is rotated on a first in,first out basis.
9. INTENSIFYING SCREEN CLEANING
-To assure that screens and cassettes are free of dust
-Suggested
Criteria:
Minimize artifacts
on films from screens or cassettes.
10. PANORAMIC SLIT ALIGNMENT TEST
-To assure that the
radiation field is in proper alignment with the slit on the panoramic
unit
image receptor.
-Suggested Criteria:
Opening on the
x-ray tube is in alignment with the opening in the image receptor.
-Steps:
1. Tape film packets to
the top and bottom of the slit on the image receptor. See
Figure 1 for set-up.
2. Mark the position of
the slit on the film packets with a ballpoint pen. (Press hard).
3. Expose the film.
4. Develop the film.
5. Observe if the
exposed portion of the film is in alignment with the demarcation
caused by the use of the
pen in Step 2.
6. Record results on the
Annual QC Checklist (Form 4)
EXAMPLE OF CHECKLIST( FORMS)
No comments:
Post a Comment